Multi Image Functional Checks; Ecg Functional Checks - GE LOGIQ Totus Basic Service Manual

Table of Contents

Advertisement

Direction 5936427, Rev. 1
4-3-20
Basic measurement functional checks (continued)
Worksheets
Measurement/Calculation worksheets are available to display and edit measurements and calculations.
There are generic worksheets as well as Application specific worksheets. The worksheets are selected
from the Measurement Touch Panel.
Report Pages
Measurements/Calculations that are included on the worksheet can also be displayed on Report Pages.
Report Pages can be customized to meet the appropriate needs of the user.
4-3-21

Multi image functional checks

1.) Press L to activate a dual screen. The single image is placed on the left side.
2.) Press R. The left side image is freezed and the image displays in the right side.
3.) Press B key to return to the single screen.
4.) Press and hold down L to activate a quad screen. The single image is placed on the upper left.
5.) Press B key to return to the single screen.
4-3-22

ECG functional checks

The ECG capability on this LOGIQ Totus is intended as a trigger for measurements, but can also be
viewed on the screen.
Parts needed: ECG cable and M5Sc-D
ECG Checks
Table 4-9
Step
1.
2.
3.
4.
5.
4 - 28
GE HealthCare CONFIDENTIAL

ECG Functional Checks

Task
Select Preset, select Cardiac, select Scan.
If you do not the ECG line, select the ECG tab, you
should selection on touch screen for DISPLAY ECG.
Select and press.
Plug in the ECG cables with nothing attached so you
pick up noise.
Connect the ECG harness to the connector on the front
of the Ultrasound system.
Connect the three leads to an ECG simulator, or fasten
the three ECG Pads to your body and connect the three
leads to respective ECG Pad.
Section 4-3 - Functional Checks
LOGIQ Totus Basic Service Manual
Expected Result(s)
You should then the line.
The line should display noise.
The LOGIQ Totus displays a straight curve along the bottom
edge of the image sector on the screen.
When connecting, the signal on the screen will be noisy.
When the connection is completed, a typical clean ECG signal
is displayed.