Direction 5936427, Rev. 1
2-2-2

Electrical requirements

General requirements
NOTE:
GE HealthCare requires a dedicated power and ground for the proper operation of its ultrasound
equipment. This dedicated power shall originate at the last distribution panel before the LOGIQ Totus.
The LOGIQ Totus will function on voltages from 100-240 volts and 50 or 60 Hz. However, if using 220
volt power, then a center tapped power source is required.
Sites with a mains power system with defined neutral and live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size ground wire from the distribution panel to the LOGIQ Totus outlet.
Sites with a mains power system without a defined neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size ground wire from the distribution panel to the LOGIQ Totus outlet.
NOTE:
Please note that image artifacts can occur, if at any time within the facility, the ground from the
main facility's incoming power source to the Ultrasound system is only a conduit.
Electrical requirements for LOGIQ Totus
In
Table
Table 2-3
PART NUMBER
DESCRIPTION
5926013/5943511
LOGIQ Totus, 100-240 VAC
Inrush current with UPS
During power on, an inrush circuit prevents the current from increasing above the stated values.
Refer to
Table 2-4
120 VAC (Console with ABCE1442-11MED)
230 VAC (Console with ABCE1442-22MED)
2 - 4
GE HealthCare CONFIDENTIAL
2-3, the electrical specifications for the LOGIQ Totus include monitor and on board peripherals.
Electrical Specifications for the LOGIQ Totus
VOLTAGE
100-240 VAC
for inrush currents at various main voltages.
Table 2-4
Inrush Current at Different Mains Voltages
VOLTAGE
115 VAC (Console only)
230 VAC (Console only)
Section 2-2 - General LOGIQ Totus Requirements
LOGIQ Totus Basic Service Manual
TOLERANCES
CONSUMPTION
+/-10%
Inrush Current
25 A Max
< 0.2 mSec
50 A Max
< 0.2 mSec
99 A Max
99 A Max
POWER
FREQUENCY
0.65 kVA
50/60 Hz
Remarks
1~4 mSec