Networking setup requirements

Stand alone scanner (without network connection)
None.
Scanner connected to hospital's network
For supported networks, a 10/100/1000 Mbps Ethernet/DICOM network connection is required.
InSite™ requirements
Need internet access on the outbound port to GE HealthCare's remote service platform (Insite), which
is only opened on request by the user and through a secure HTTPS connection on port 443.
Insite requires an Ethernet connection with a 10/100 Mbps or 10/100/1000 Mbps interface.
Purpose of the DICOM network function
DICOM services provide the operator with clinically useful features for moving images and patient
information over a hospital network. Inbound and outbound connections through port 104 are for
DICOM connectivity. Examples of DICOM services include the transfer of images to workstations for
viewing or transferring images to remote printers. As an added benefit, transferring images in this
manner frees up the on-board monitor and peripherals, enabling viewing to be done while scanning
continues. With DICOM, images can be archived, stored, and retrieved faster, easier, and at a lower
cost.
DICOM option setup requirements
To configure the LOGIQ Totus to work with other network connections, the site's network administrator
must provide information to complete the form in
Information" on page
Entries must include:
A host name, local port number, AE Title, IP address and Net Mask for the LOGIQ Totus.
The IP addresses for the default gateway and other routers at the site for ROUTING
INFORMATION.
The host name, IP address, port and AE Title for each device the site wants connected to the
LOGIQ Totus for DICOM APPLICATION INFORMATION. A field for the make (manufacturer) and
the revision of the device, is also included. This information may be useful for troubleshooting.
For connectivity setup information the current revision of the LOGIQ Totus Basic User Manual (latest
revision).
GE HealthCare CONFIDENTIAL
2-14. Ensure that there are no spaces in any field of the form.
1-4-1 "Label Location" on page
Chapter 2 Site Preparations
Figure 2-3 "Worksheet for DICOM Network
1-13.
LOGIQ Totus Basic Service Manual
2 - 13