GERAFT
D
5604324, R
IRECTION
EVISION
2-3-4
Networking Setup Requirements
2-3-4-1
Stand-alone Unit (without Network Connection)
None
2-3-4-2
Unit Connected to Hospital's Network
Supported networks:
•
Ethernet network connection
•
Up to 900 Mbps WLAN (optional)
2-3-4-3
Purpose of the DICOM Network Function
DICOM (Digital Imaging and Communications in Medicine) services provide the operator with clinically
useful features for moving images and patient information over a hospital network.
Examples of DICOM services include the transfer of images to workstations for viewing or transferring
images to remote printers. As an added benefit, transferring images in this manner frees up the on-
board monitor and peripherals, enabling viewing to be done while scanning continues.
With DICOM, images can be archived, stored, and retrieved faster, easier, and at a lower cost.
2-3-4-4
DICOM Option Pre-installation Requirements
To configure the LOGIQ P9/P7 ultrasound unit to work with other network connections, the network
administrator must provide some necessary information.
Use the
• LOGIQ P9/P7
Details:
• Routing Information:
• DICOM Application Information: Details of DICOM devices in use at the site, including the DICOM
Installation see:
2-10
11
DRAFT (J
Agent Configuration (R3 only) on page 3-65
Section 3-13 "Configuring Connectivity" on page
Section 2-3 - Facility Needs
24, 2019)
ANUARY
to record required information that must include:
DICOM network details for the LOGIQ P9/P7 unit, including the
host name, local port, IP address, AE title and net mask.
IP addresses for default gateway and other routers in use at site.
host name, AE title, DICOM port number and IP addresses.
LOGIQ P9/P7
S
ERVICE
3-58.
M
ANUAL