Pulse Doppler Mode

1) From Utility -> Imaging -> General, verify check on "Simultaneous".
2) Press B Mode button.
3) Press PW Mode button. (Active PW Mode button on "User Configurable Mode Rotary".)
4) Verify that system scan B/PW duplex with the activated probe.
5) Revert to original setting in Utility -> Imaging -> General "Simultaneous".
4-3-7-5

B/CF/PW Mode

1) Press B Mode button.
2) Press PW Mode Button.
3) Press CF Mode Button.