Replacement Of The T2 Opio Base Rear Cover Assy - GE LOGIQ Totus Basic Service Manual

Table of Contents

Advertisement

D
5936427, R
IRECTION
EV
8-7-7

Replacement of the T2 OPIO Base Rear Cover Assy

This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
Table 8-43
Manpower
One person
8-7-7-1
Preparations
1.) Remove the OPIO Rear Tray Set-T as described in
Set-T" on page 8-187
2.) Remove the Probe Holder Set-T as described in
page 8-90
8-7-7-2
Removal Procedure
Step
Unscrew 5 screws to remove the T2 OPIO Base
1.
Rear Cover Assy from the console.
8-7-7-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform "Functional Checks after FRU replacement (Debrief Guideline)" as described in
12 on page 8-223
GE HealthCare CONFIDENTIAL
. 1
Manpower, Time, Tools, and PPE
Time
Approximately 5 minutes
Section 8-7 - Replacing OPIO
LOGIQ T
Tools
Philips Screwdriver
8-11-4 "Replacement of the OPIO Rear Tray
8-7-3 "Replacement of the Probe Holder Set-T" on
Corresponding graphic
Basic Service Manual
OTUS
8-2-6 "Tools needed
PPE
Cut resistant gloves
Knee Pad
Section 8-
8 - 95