Rear Cover & Latch Assembly Replacement Procedure
8-2-4-1
Tools
None required.
8-2-4-2
Time Required
30 minutes
8-2-4-3
Preparations
Shut down the Vivid i n ultrasound unit, as described in
8-2-4-4
Rear Cover & Latch Assembly Removal Procedure
1) Disconnect the AC adapter and all cables from the Vivid i n.
2) Remove the Control Panel and Keyboard as described in the
Removal Procedure" on page
3) Remove the LCD Display Frame as described in the
page
4) Remove the Microphone Assembly as described in the
Procedure" on page
5) Remove the LCD Display; see the
6.) Loosen and remove the two screws on each of the left and right hinges (attached to the Rear Cover
and Latch Assembly), as shown in
7.) Lift the Rear Cover and Latch Assembly away from the Bottom Assembly and remove completely
from the system.