Microphone Assembly Installation Procedure

1) Using the replacement Microphone Assembly, re-connect the miniature plug to the cable connector
below the LCD Display (refer to Figure 8-54 on page 8-54 ).
2) Feed the microphone cable into the cable groove that runs along the side and bottom of the LCD
Display.
Note:
An adhesive rubber channel is fitted inside the groove to hold the cable in place.
3) Clip the microphone into position, making sure it is properly seated on the upper right side of the
LCD Display (refer to Figure 8-53 on page 8-54 ).
4) Return the LCD Display Frame as described in the
page
5.) Proceed to perform the following functionality tests:
-
-
5
EVISION
8-23.
Verify that the microphone is functioning.
"Chassis Current Leakage Test" on page 10-25