7-3-2-2

DC Offset Verification

NOTE:
When performing this test, no probe should be connected to the system and nothing should be touching
the probe connectors!
1.) In the Diagnostic Test window, trackball to DC Offset Verification in the list of tests available for
selection from the Diagnostic Test Tree (see
Verification name is displayed under Execute; the relevant system components to be tested are
highlighted light blue in the Schematic View (PM, TR32 and RFI boards), as shown below:
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
5.) Trackball to the Start button and press Set.
7-18
5
EVISION
Figure 7-8 DC Offset Verification Test
Section 7-3 - Performing Full System Diagnostics
Figure 7-8 below
) and press Set. The DC Offset
V
S
M
IVID I N
ERVICE
ANUAL