Verifying The Transportation Box Contents; Physical Inspection; System Voltage Settings; Emi Protection - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FL091021, R
IRECTION
3-4-3

Verifying the Transportation box Contents

After unpacking, it is important to verify that all items ordered by the customer have been received (refer
to "Verifying the Transportation Box Contents" on page 3-12). Compare all items listed on the packing
slip (shipping consignment note) with those received and report any items that are missing, back-
ordered, or damaged, to your GE Healthcare sales representative. The contact address is shown in
1-8-1 "Contact Information" on page
It is recommended to keep and store the Transportation Box and all other packing materials in case the
NOTICE
unit has to be moved to a different location in the future.
For warranty purposes, storage of the above is required for one year from date of purchase.
3-4-4

Physical Inspection

3-4-4-1

System Voltage Settings

Verify that the Vivid i n ultrasound AC adapter and Cart (if applicable) are set to the correct voltage.
The Voltage settings are 220-240V AC - (Europe, Latin America, and China).
WARNING Setting the Vivid i n ultrasound components to the wrong voltage setting will most
likely destroy the equipment.
3-4-5

EMI Protection

The Vivid i n Ultrasound Unit has been designed to minimize the effects of Electro Magnetic
Interference (EMI). Many of the covers, shields, and screws are provided primarily to protect the system
from image artifacts caused by this interference. For this reason, it is imperative that all covers and
hardware are installed and secured before the unit is put into operation.