Rechargeable Battery Pack Assy (GPA)

The lithium ion rechargeable battery provides a backup mechanism to the Vivid i n system whenever
an AC power source is not available. The battery module is supplied with a lithium ion rechargeable
battery pack (GPA) installed in the battery bay, as standard.
The Vivid i n scanner has built-in charger functionality (see Figure 5-23 on page 5-29 ) and switches
automatically from battery operation to AC operation and vice versa.
When shutting down the system, leave the main power cable connected to keep the battery fully charged.
NOTE:
The lithium ion technology used in the system's battery is significantly less hazardous to the
environment than the lithium metal technology used in some other batteries.
NOTE:
The battery is designed to be replaced every 2 years.
CAUTION: The battery is designed to work with Vivid i n systems only. Only use the batteries
authorized by GE Healthcare.
WARNING: Used batteries should not be placed with common household waste products. Contact your
local authority for the location of a chemical waste collection program nearest to you.
CAUTION: The battery has a safety device.
-
-
-
-
-
-
-
-
-
CAUTION: If the Vivid i n portable ultrasound system is not being used for periods of one month or
longer, the battery should be removed during lengthy non-use period.
CAUTION: To avoid the battery bursting, igniting, or fumes from the battery causing equipment
damage, observe the following precautions:
-
-
-
-
5
EVISION
Do not disassemble or alter it. Charge the batteries only when the ambient temperature is
between 0° and 65° C (32° and 149° F) and discharge the batteries between -10° and 55° C
(14° and 131° F).
Do not short-circuit the battery by directly connecting the battery terminals with metal objects.
Do not heat the battery or incinerate.
Do not expose the battery to temperature over 60° C (140° F). Keep it away from fire and
other heat sources.
Do not charge the battery near a heat source, e.g. fire or heaters.
Do not leave the battery in direct sunlight.
Do not pierce the battery with a sharp object, hit it, or step on it.
Do not use a damaged battery. Do not solder a battery.
Do not connect the battery to an electrical outlet.
Do not immerse the battery in water or allow it to get wet.
Do not put the battery into a microwave oven or pressurized container. If the battery leaks or
emits an odor, remove it from all possible flammable sources.
If the battery emits an odor or heat, is deformed or discolored, or in a way appears abnormal
during use, recharging or storage, immediately remove it and stop using it.
If you have any questions about the battery, consult GE or your local representative.