Functional Checks; Basic Controls; Alphanumeric Keyboard Test; Alphanumeric Keyboard Test - Recommended - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FL091021, R
IRECTION
EVISION
Section 4-3

Functional Checks

4-3-1

Basic Controls

4-3-1-1

Alphanumeric Keyboard Test

Table 4-5
Step
1
2
3
4
5
4-3-1-2
External Keyboard Test
NOTE:
The External Keyboard test also includes testing of the following components: TGC Sliders, Rotary,
Keyboard backlight, Button backlight, and Active Button.
Table 4-6
Step
1
2
3
4
5
6
7
5

Alphanumeric Keyboard Test - Recommended

Task
Power-up the System in Normal Scanning Mode
Press Config and select Report
Select the Comment Text Tab
Press all Alphanumeric keys one-by-one.
To quickly test every key (each letter in the alphabet), type
the following sentence:
The quick brown fox jumps over the lazy dog
Repeat the test, while holding down the Shift key.
Use the Backspace key to delete the text you typed.
External Keyboard Test Vivid i n- Recommended 1 of 3
Task
Power-up the System in Normal Scanning Mode
Press <Alt> F2 to start the diagnostic package.
Highlight the Keyboard Test and press <Start>
Trackball to Connect and press Set.
Press all extended KB keys one-by-one.
Press the Clear soft button to clear the Event window.
Perform the TGC Slider Test, as follows:
a) Move all TGC sliders to the left position
b) Move all TGC sliders to the right position
c) Move all TGC sliders to the central position
Perform the ROTARY TEST, as follows:
Rotate the rotary (in both directions) and press the button
at the center of the rotary.
Chapter 4 - General Procedures and Functional Checks
V
IVID I N
Expected Result(s)
As each key is pressed, the corresponding letter is
typed on the screen.
Expected Result(s)
The keyboard Test window opens
The extended KB keys activate properly. Every hit on a key
is followed by two rows (Depress/Release).
The position value is 0 (+2) for all TGC Sliders
The position value is 63(-2) for all TGC Sliders
The position value is 31(±2) for all TGC Sliders
The rotary is activated properly; pressing on the button will
display a row of data for pressing and another row of data
for releasing (Depress/Release).
S
M
ERVICE
ANUAL
4-17