Electrical Requirements

NOTE:
GE Medical Systems requires a dedicated power and ground for the proper operation of its
Ultrasound equipment. This dedicated power shall originate at the last distribution panel before
the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size ground wire from the distribution panel to the Ultrasound outlet.
NOTE:
Please note that image artifacts can occur, if at any time within the facility, the ground from the
main facility's incoming power source to the Ultrasound unit is only a conduit.
2-2-2-1
LOGIQ™ 5 Power Requirements
The following power line parameters should be monitored for one week before installation. We
recommend that you use an analyzer Dranetz Model 606-3 or Dranetz Model 626:
Table 2-11
Voltage Range
Power
Line Frequency
Power Transients
Decaying Oscillation
2-2-2-2
Inrush Current
Inrush Current is not a factor to consider due to the inrush current limiting properties of the power
supplies.
2-2-2-3
Site Circuit Breaker
It is recommended that the branch circuit breaker for the machine be ready accessible.
POWER OUTAGE MAY OCCURE.
CAUTION
The LOGIQ™ 5 requires a dedicated single branch circuit. To avoid circuit overload and possible
loss of critical care equipment, make sure you DO NOT have any other equipment operating on
the same circuit.
1
Electrical Specifications for LOGIQ™ 5
PARAMETER
100V
220V
115V
All applications
All applications
All applications
All applications
Chapter 2 Pre-Installation
AREA
100 VAC ±10% (90-110 VAC)
220-240 VAC ±10% (198-264 VAC)
115 VAC ±10% (103-127 VAC)
MAX. 1250 VA
50/60Hz (±2Hz)
Less than 25% of nominal peak voltage
for less than 1 millisecond for any type of
transient, including line frequency,
synchronous, asynchronous, or
aperiodic transients.
Less than 15% of peak voltage for less
than 1 millisecond.
LOGIQ™ 5 S
M
ERVICE
ANUAL
LIMITS
2 - 3