Measurement And Multi Image Checks; Performing B Mode Measurements Checks - GE LOGIQ A3 Basic Service Manual

Table of Contents

Advertisement

GE H
EALTHCARE
D
5213326-100, R
IRECTION
4-4-4

Measurement and Multi Image Checks

4-4-4-1

Performing B Mode Measurements Checks

Step
1.
2.
3.
4
5.
6.
4-8
Approv e d Docu m ent - 5213326-100TPH_r3.pdf Page 95 of 252
3
EVISION
Figure 4-3 Measurement Keys
Table 4-8
Measurement Checks
Task
Generate the B image.
Press Freeze
Press MEASURE on the Control Panel to
enter the Assign and Measure modality.
Trackball the cursor
Press Set.
Trackball the cursor
Section 4-4 - Functional Checks
LOGIQ™ A3 S
Expected Result(s)
B image is active on the screen
Freeze the image to measure.
The measurement soft menu for the current
exam category is displayed on the Monitor.
Select the appropriate exam category and
measurement kind.
Move the cursor to the start point of the
measurement
Anchor the start point of the measurement.
Move the cursor to the measurement end
point.The current distance value is displayed
in the Measurement result table and is
instantaneously updated when moving the
cursor
M
ERVICE
ANUAL