Safety Reminders - GE LOGIQ A3 Basic Service Manual

Table of Contents

Advertisement

GE H
EALTHCARE
D
5213326-100, R
IRECTION
3-2-1

Safety Reminders

WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
DANGER
AC GROUND LINE (I.E., METER'S GROUND SWITCH IS OPEN), DON'T TOUCH
THE UNIT!
If the unit is very cold or hot, do not turn on its power until it has had a chance to acclimate
to its operating environment.
CAUTION
To prevent electrical shock, connect the unit to a properly grounded power outlet. Do not use
a three to two prong adapter. This defeats safety grounding.
CAUTION
Do NOT wear the ESD wrist strap when you work on live circuits and more than 30 V peak is
present.
CAUTION
Do not operate this unit unless all board covers and frame panels are securely in place.
System performance and cooling require this.
CAUTION
OPERATOR MANUAL(S)
The User Manual(s) should be fully read and understood before operating the LOGIQ A3 and
CAUTION
kept near the unit for quick reference.
ACOUSTIC OUTPUT HAZARD
Although the ultrasound energy transmitted from the LOGIQ A3 probe is within AIUM/NEMA
CAUTION
standards, avoid unnecessary exposure. Ultrasound energy can produce heat and
mechanical damage.
NOTE:
Check the shipping container for special instructions. Verify that the container is intact. In some cases
a secondary container may be used. If so, ask the carrier for unpacking instructions.
Approv e d Docu m ent - 5213326-100TPH_r3.pdf Page 66 of 252
3
EVISION
Figure 3-6 LABELS ON PACKAGE
Chapter 3
LOGIQ™ A3 S
M
ERVICE
ANUAL
3-9