Troubleshooting Tips - GE LOGIQ A3 Basic Service Manual

Table of Contents

Advertisement

GE H
EALTHCARE
D
5213326-100, R
IRECTION
Section 7-4Troubleshooting Tips
S.NO
1
2
3
No Image on External Monitor
4
5
6
Peripherals Power Supply from System
7
8
9
10
11
12
13
7-8
Approv e d Docu m ent - 5213326-100TPH_r3.pdf Page 127 of 252
3
EVISION
Issue
System doesn't Bootup
No Display
No Keyboard Response
Probe Not Recognised
No Image
Absent
Unable to Print
Unable to archive/backup
Unable to Freeze
TGC Doesn't Work
Gain Doesn't change
Trackball doesn't Work
Section 7-4 - Troubleshooting Tips
Possible Cause
Circuit Breaker Trip
Switch On Circuit Breaker
Transformer Failure
Replace Transformer
HvLv Failure
Replace HvLv
MST Failure
Replace MST
Circuit Breaker Trip
Switch On Circuit Breaker
Transformer Failure
Replace Transformer
HvLv Failure
Replace HvLv
LCD Failure
Replace LCD
Loose connection at Back
Restore Loose Connections
Plane Board
MST Failure
Replace MST
Keyboard Failure
Replace Keyboard
Loose connection at Back
Restore Loose Connections
Plane Board
MST Failure
Replace MST
MST Failure
Replace MST
Probe Loose Connection
Restore Loose Connections
CNB Failure
Replace CNB
MST Failure
Replace MST
CNB Failure
Replace CNB
TxRx Failure
Replace TxRx
DIF Failure
Replace DIF
MST Failure
Replace MST
Fuse Failure/Transformer
Replace Transformer
Failure
MST Failure
Replace MST
MST Failure
Replace MST
KBD Failure
Replace KBD
MST Failure
Replace MST
TGC Failure
Replace TGC
MST Failure
Replace MST
KBD Failure
Replace KBD
MST Failure
Replace MST
KBD Failure
Replace KBD
MST Failure
Replace MST
LOGIQ™ A3 S
M
ERVICE
ANUAL
Resolution

Advertisement