Select the connected sensor from the available sensors with the
and
key, and press EXE. Press CE to leave menu.
The next menu gives you all the details of the selected sensor. This
information should be checked with the sensor information on the
connector. If both sets of information do not match, proceeding may
cause serious problems. In the case of a mismatch press CE to
leave. Providing that the information matches, press EXE to confirm.
Choose direct or scale to open the working screen.
The working screen shows the pressure measurement. Also during
working mode, the serial number of the connected sensor is shown in
the display.
Pressure Switch
Make a calibration hook-up as indicated and switch the calibrator on.
Select SWITCH in the measure menu and PRESSURE in the source
menu. Select the applicable sensor to open the working screen.
Press # to unlock the switch display. Increase or decrease the
calibration pressure. The switch reading tracks the source reading
until the switch position changes. This change will immediately
"freeze" the switch (upper) reading indicating the actual switch
setting.
EN-20