Safety
Service and Maintenance
2.3.1 Operation, Installation and Repair
WARNING!
To avoid risk of electric shock, this equipment must only be connected to a supply
mains with protective earth.
WARNING!
Only medical-approved products shall be in the X-ray room
Risk of electric shock to patient or user
• No non-medical electrical devices shall be used in the x-ray room.
• Note that the monitor and the PC for the Image system, are none-medical approved
products.
WARNING!
The Manufacturer can not assume responsibility for the safety features or for the reli-
ability and performance of the equipment, if:
• installation of equipment expansions or modification are not approved by the
Manufacturer.
• installation of equipment expansions or modification are not carried out by persons
authorized by the Manufacturer.
• components are not replaced by original spare parts in case of a malfunction.
• the electrical installation of the room concerned does not meet the requirements or
the corresponding national regulations.
• the product is not used in accordance with the operating instructions.
CAUTION!
Only service engineers are allowed to open the covers.
CAUTION!
When installing this equipment in a different location, contact us or our designated
dealer.
CAUTION!
Do not modify the equipment.
CAUTION!
Do not remove, disassemble, change, modify, repair, or add any part.
OMNERA®
0073-095-418-EN Rev. 5.1
Page 12
Operator´s Manual
2020-09-25