User Interface
Ceiling Suspended x-ray Tube Support

3.4.2 Servo Button

When pressing the servo button, the System will enter an active mode status. The Indicator
light, will show the status of the mode.
If the servo button is activated, the stand will start positioning automatically (excluding Free
mode). If the servo button is released, the movement will stop and a manual movement of
the stand is possible.
If a new position is chosen or if the OTC is manually moved from the position, the OTC will
automatically start moving when the servo button is activated.
3.4.2.1 External Servo Button
When pressing the external servo button, the System will enter an active mode status.
The indicator light, will show the mode status.
The stand will start positioning automatically (excluding Free mode).
If the external servo button is released, the movement will stop and manual movement of the
stand is possible.
If a new position is chosen or if the OTC is manually moved from the position, the OTC will
automatically start moving if the external servo button is activated.
The external servo button is also used for auto positioning the overhead tube crane. When
the servo button is released the movement stops. The servo button includes an indication
light, see description below. See Fig. 3-30 External servo button.
The external servo button is equipped with an emergency stop, that can be used to stop the
motions in the system.
1. Emergency stop
2. External Servo button and Indicator light