5.1 Point of use (Pre-cleaning)
Pre-cleaning should be done immediately after each
use. The procedure is as follows:
1) Remove contaminated needle guide and probe cover from the probe with
needle guide bracket and dispose them appropriately.
2) Remove the needle guide bracket from the probe.
3) Flush patient's blood or fluid by tap water from the needle guide
bracket directly after use, until the surface looks visually clean
4) Immerse the needle guide bracket in sufficient amount of high quality
tap water. Scrub it using soft cloth to remove all visible soil and
from its surface.
5) Prepare following items before manual pre-cleaning of the puncture
adapter using ultrasound:
a) Ultrasound bath: i.e. Euronda, Eurosonic 4D
b) Detergent: Cidezyme® (Johnson & Johnson, #2258) or another
cleaning agent with approved material compatibility for this
medical device
c) Personal
protective skirt, face protection mask or protective glasses,
see also instructions of the manufacturer for the detergent and
the disinfectant)
Pre-Cleaning:
1) Fill the ultrasound bath with the detergent. The temperature of
the detergent solution should be between 15-30 °C, concentration
is
1.6%.
manufacturer's instructions. Please also consider the material
compatibility for the medical device.
2) Immerge the needle guide bracket into the diluted detergent in
the ultrasound bath and switch on the ultrasound bath for a time
of 20 minutes at 25°C.
3) Remove the needle guide bracket and continue with the manual or

automated cleaning and disinfection.

5.2 Containment and transportation
Putting the contaminated equipment into exclusive
shock and damage proof container for transportation
is recommended. It is recommended that instruments are reprocessed as
soon as possible and not later than 4 hours after usage.
5.3 Manual Cleaning and disinfection
protective
equipment
If
a
different
(gloves,
detergent
-31-
Point of use
(Pre-cleaning)
water
repellent
is
used,
follow
Containment and
transportation
Q1E-EP1457
the