4) Visually check the outer surface of the Magnetic sensor attachment
and the Spacer for EZU-RV2S for leavings of the disinfectant. If
necessary, repeat the rinsing.
Water, Detergent or
Disinfectant
Fig.11 Immersion of the Probe, the Magnetic sensor attachment
and the Spacer for EZU-RV2S
5.4

Automated cleaning and disinfection

The following items must be provided prior
to automated cleaning and disinfection:
a) Washer disinfector: according to DIN
EN ISO 15883 with chemo-thermal
program (temperature: max 60° C).
b) Waterproof box EZU-WB1
c) Detergent: Korsolex Endo-Cleaner (Bode Chemie; # 972 020) or another
cleaning agent with approved material compatibility for this medical
device
d) Disinfectant: Korsolex Endo Disinfectant (Bode Chemie; # 972 030)
or another disinfectant with approved material compatibility for
this medical device
Washer Disinfector
Automated Cleaning
Automated Disinfection
- 19 -