Drying - Hitachi EUP-B512 Instruction Manual

Biopsy probe
Table of Contents

Advertisement

1) The parameters of the cleaning and disinfection of the device are
as follows:
Program step
Pre-Rinse
Cleaning
Deionized water
Rinse
Deionized water
Disinfection Deionized water
Rinse
Deionized water
Rinse
Deionized water

Drying

2) Connect the waterproof box EZU-WB1 to the probe connector and use
the tester to confirm that there is no air leak.
Refer to the instruction manual of the waterproof box EZU-WB1 for
detail information.
3) Place the probe in the baskets of the washer disinfector.
4) Close the door of the washer disinfector and start the chemo-thermal
program.
5) Open the door after the process is done.
Take the probe out of the washer disinfector and check that it is dry.
If not, dry them as described in the chapter drying.
5.5
Drying
1) Wipe the equipment with a single-use, fluff-free wipe or towel to
remove moisture from the surface of the equipment.
2) Dry the equipment naturally in an ambient temperature between 15-30°C
for a minimum of 4 hours. Alternatively the equipment can be dried
using a drying heater at a temperature of less than 60°C.
Water (40 l)
Dosage (ml/l) Temp. (°C) Time (min)
Cold water
-
-
5 (0.5%)
-
10 (1%)
-
-
-
- 20 -
-
5
50
10
-
1
55
5
-
1
55
1
55
15
Drying
Q1E-EP0780