Point Of Use (Pre-Cleaning); Containment And Transportation - Hitachi EUP-B512 Instruction Manual

Biopsy probe
Table of Contents

Advertisement

5.1

Point of use (Pre-cleaning)

Pre-cleaning should be done immediately after each
use. The procedure is as follows:
A) EUP-B512
1) Remove the protective cover.
2) Clean the probe of all patient's blood or fluid with running tap
water until the surface of the probe looks visually clean.
3) Wipe the whole surface of the probe with gauze pad and remove
superficial visible impurities.
B) Magnetic sensor attachment and Spacer for EZU-RV2S
1) Remove the Magnetic sensor attachment and the magnetic sensor from
the probe. Remove the Spacer for EZU-RV2S from the magnetic sensor
if the sensor is EZU-RV2S.
2) Immerse the Magnetic sensor attachment and the Spacer for EZU-RV2S
in sufficient amount of high quality tap water. Scrub them using
soft cloth to remove all visible soil from their surface.
5.2

Containment and transportation

Putting the contaminated equipment into exclusive
shock and damage proof container for transportation
is recommended. It is recommended that instruments
are reprocessed as soon as possible and not later than 4 hours after
usage.
- 15 -
Point of use
(Pre-cleaning)
Containment and
transportation
Q1E-EP0780