Manual Cleaning And Disinfection - Hitachi EUP-B512 Instruction Manual

Biopsy probe
Table of Contents

Advertisement

5.3

Manual Cleaning and disinfection

Prepare following items before manual
cleaning and disinfection:
a) Detergent: Cidezyme (Johnson &
Johnson, #2258) or another cleaning
agent with approved material
compatibility for this medical device
b) Disinfectant: Cidex OPA (Johnson &
Johnson, # 20391) or another
disinfectant with approved material
compatibility for this medical device
c) Two tanks, one for cleaning and one for
disinfection - optional:
1 additional tank for rinsing with
deionized/tap water (sufficient size
for immersion of the submergible part of the probe at full length)
d) Soft, fluff free cloth or single use towel
e) Personal protective equipment (gloves, water repellent protective
skirt, face protection mask or protective glasses, see also
instructions of the manufacturer for the detergent and the
disinfectant)
Manual Cleaning:
Prepare the detergent solution in a tank with cold water (please follow
the instructions of the detergent manufacturer regarding application,
dilution and contact time).
A) EUP-B512
1)
The temperature of the detergent solution should be between 15-30 °C,
concentration is 1.6%. Please note the minimum contact time of the
detergent in the manufacturer's instruction. If a differing
detergent is used, please also note the approved material
compatibility for the medical device.
2)
Immerse the immersible part of the probe without connector into
the diluted detergent solution (see Figure 11). Wipe the immersible
part of the probe under the surface of the detergent solution with
a soft cloth to remove all visible soil. Be sure that all grooves
of the probe are implemented during the cleaning process.
3)
The immersible part of the probe should be left in the detergent
solution according to the specified contact time of the detergent
manufacturer.
Manual cleaning and
Rinsing after manual
Rinsing after manual
- 16 -
disinfection
Manual Cleaning
cleaning
Manual Disinfection
disinfection
Q1E-EP0780