Sterilization - Hitachi EUP-B512 Instruction Manual

Biopsy probe
Table of Contents

Advertisement

5.8

Sterilization

The probe and accessory can be sterilized using either ethylen oxide
gas (EtO) sterilization or plasma sterilization (see table below).
Follow the manufacturer's instructions of the sterilizer regarding
usage, temperature and sterilization-time.
The sterilization method and operating conditions are as follows.
Sterilization Method
Plasma Sterilization:
STERRAD® 50, 100S or 200 (*)
Plasma Sterilization:
Sterrad® NX or 100NX (*)
ETO Sterilization
* STERRAD® systems are manufactured by "Johnson & Johnson"
1) Before performing sterilization, check that the operation data of
sterilizer are in conjunction with min. and max. data applicable
for the probe, the Magnetic sensor attachment and the Spacer for
EZU-RV2S.
2) Do not sterilize the probe, the Magnetic sensor attachment and
the Spacer for EZU-RV2S by Steam Autoclaving. If you autoclave
them, they suffers serious damage and will be not functional.
Condition
Short Cycle
Standard cycle
Gas Type: 10% EO/ 90% HCFC
Temperature: 50-55°C
Exposure Time: More than 120 minutes
Pressurization: 162-200kPa
Depressurization: 13-8kPa
Relative humidity: 40-90%
Aeration is minimum 12 hours
WARNING
- 22 -
Sterilization
Q1E-EP0780