Air Flow Control; Air Flow Distribution; Filters; Air Flow Inside The Scanner - GE LOGIQ 7 Service Manual

Hide thumbs Also See for LOGIQ 7:
Table of Contents

Advertisement

GE H
EALTHCARE
D
2286865, R
IRECTION
EVISION
Section 5-7

Air Flow Control

5-7-1

Air Flow Distribution

A
B
A
B
The four air flow pathes allow the scanner to be cooled down as shown in the figure above.
Path A (Front lower left > Filter > PC Box > Rear upper left) for PC Box cooling.
Path B (Front lower left > Filter > LV unit > Rear lower left) for LV unit cooling.
Path C (Bottom right > Filter > NEST Assy > Rear upper right) for NEST Assy cooling.
Path D (Rear bottom right > Filter > HV unit > Rear upper right) for HV unit cooling.
5-7-2

Filters

The scanner contains the three filters located at:
-
-
-
5-42
14
Figure 5-67 Air Flow Inside the Scanner
Front lower left for air flow of the LV unit and PC box.
Bottom right for air flow of the NEST Assy.
Rear bottom right for air flow of the HV unit.
Section 5-7 - Air Flow Control
LOGIQ™ 7 S
M
ERVICE
ANUAL
C
D
C
D

Advertisement

Table of Contents
loading

Table of Contents