Air Flow Control - GE LOGIQ E10 Basic Service Manual

Hide thumbs Also See for LOGIQ E10:
Table of Contents

Advertisement

D
5750007-1EN, R
IRECTION
Section 5-11

Air Flow control

5-11-1
General description
The Air Flow Control includes the following components: Air cooling fans located at the bottom of the
LOGIQ E10 Card Cage and air cooling fans in the power supply that are self-managed by the power
supply itself.
5-11-2
Location in the LOGIQ E10
There are three air strainers/filters: One located on the rear of the LOGIQ E10 and one located on each
side panel, near the bottom in a recessed feature. Refer to Chapter 8 for more information.
Figure 5-6 Air Flow Path of the LOGIQ E10
Cooling air is drawn in, evenly into the filters located in the bottom of the Side Covers/Air Intake and the
Rear Bumper/Air Intake. The warmed air is exhausted from the Rear Cover/Air Exhaust.
. 1
EV
Chapter 5 Components and Functions (Theory)
LOGIQ E10 B
S
ASIC
ERVICE
M
ANUAL
5 - 25

Advertisement

Table of Contents
loading

This manual is also suitable for:

5935000e10

Table of Contents