Section 5-13

Air Flow Control

5-13-1

General Description

The Air Flow Control includes the following components:
Dust Filter
Dust Filter Frame
Air Inlet Channel
Fan Assembly
Other components that influence on the airflow is the Front-End Board Rack itself. The air is sucked in
from the rear side of the system through the Filter Cover and the Filter. It passes the Fan Assembly
directly into the upper part of the rear side of the Front-End Board Rack and makes a cooling air current
between the Boards before it blows out against the floor. See
5-13-2

Location in the Unit

The Fan Assy is located between the Board Rack and the Filter, which can be seen on the rear side of
the System.
AIRFLOW THROUGH BOARD
RACK BEHIND SIDE COVER
AIR OUTLET UNDER THE SYSTEM
5 - 146
11
Figure 5-99 Airflow through the System
Section 5-13 - Air Flow Control
V
7 S
IVID
ERVICE
Figure
5-99.
AIR INLET THROUGH FILTER
FAN ASSY
M
ANUAL