6-4-2
later and LS6 17 inch IPS-Pro type LCD Upgrade systems)
6-4-2-1
Adjusting Parameters
There are 4 parameters to control the brightness of the IPS-Pro type LCD monitor. This section
describes the property and adjustment for each parameter.t.
6-4-2-1-1
Brightness
Range: 0 <-> 100
Property: Darker <-> Brighter.
Factory default: 80
Adjustment: Change the value by directly pressing the button (2) and (3), and press the button (1)
to exit.
6-4-2-1-2
Contrast
Range: 0 <-> 100
Property: Less Contrast <-> More Contrast.
Factory default: 100
Adjustment: Change the value in Advanced Menu according to the procedure below.
Go into the Contrast Menu
6-12
3
IPS-Pro type LCD Monitor (5261237) Brightness Adjustment (BT08 or
Figure 6-89 17 inch IPS-Pro Type LCD Monitor
Adjustment buttons
1.) Press and hold the button (1) for 15 seconds, and Advanced Menu appears.
2.) Press the button (3) once to select the Color menu ( [RGB] icon ).
3.) Press the button (1) to go into Color menu. The upper left end icon is highlighted in the
menu.
4.) Press the button (1) to go into the Brightness and Contrast menu.
5.) Press the button (1) to select Contrast menu