Lcd Monitor Position Adjustment; Logging On To Vivid T8/Vivid T8 Pro As "Adm - GE Vivid T8 Basic Service Manual

Hide thumbs Also See for Vivid T8:
Table of Contents

Advertisement

4-2-3 Power off (continued)
4-2-3-1 System shutdown
CAUTION

4-2-4 LCD Monitor position adjustment

4-2-5 Logging on to Vivid T8/Vivid T8 Pro as "ADM"

Vivid T8/Vivid T8 Pro – Basic Service Manual
5718826-100 English Rev. 10
Disconnect the Mains Power Cable is necessary. For example:
Relocating the scanner.
DO NOT unplug and/or transport the unit until after the power
off sequence has been completed. Failure to do so may result
in corrupted patient files.
Refer to User Manual for LCD Monitor position adjustment.
Select Utility tab on the touch screen, then select Config.
It will bring up the Operator Login dialog where you must log
on.
Figure 4-2. Operator Login Window
1. Operator: Select the operator.
2. Password: Enter Operator's password (optional).
3. Select the type of Login or Cancel.
Emergency: Stores data only from current patient
examination.
OK: Standard login.
Cancel: Cancel login.
As default, two users are defined, USR and ADM.
USR
If you log on as USR, you will have access to do set-up tasks
that a user may need to do during daily use.
General procedures
4-9