Panasonic MCO-170AICL Operating Instructions Manual page 48

Co2 incubator

Advertisement

4. On the Alarm screen, the newest 7 days' alarm
logs (containing that day) are displayed.
Note: When the number of applicable alarm log is 6
or more, by pressing the top (▲) or the bottom (▼)
log, the log table currently displayed scrolls and
hidden alarm logs can be seen.
・Press Back key to return to the Log screen.
・Press Top key to return to the Top screen.
5. On the Alarm screen, by inputting days into the
Last XX Days input box, alarm logs for specified
days (containing that day) are displayed.
Settable range: 1 day~45 days.
Note: The error of about 1 minute may be observed
during 1 month. Refer to page 51 for the procedure
of setting time.
・Press Back key to return to the Log screen.
・Press Top key to return to the Top screen.
●On the Alarm screen of procedure 4 or 5, alarm log data can be exported in CSV format to the USB
memory inserted into the USB port.
6. Insert the USB memory into the USB port.
Note: It is not possible to use a USB memory with security functions that requires entering password.
7. Press Export key.
8. When the export is complete, Information dialog
box is displayed. Press OK key. Refer to page 49
and 50 for the details about abnormal export or
exported file name.
9. Press Top key to return to the Top screen.
47