Download Print this page

Performance - Panasonic MCO-170AICL Operating Instructions Manual

Co2 incubator

Advertisement

PERFORMANCE

Product name
Model No.
Temperature control range
Temperature distribution
Temperature variation
CO
control range
2
CO
variation
2
Chamber humidity
Applicable environment
condition
Noise level
Power consumption
Heat emission
Rated voltage, frequency
Amperage
*When ambient temperature is 25
temperature, the maximum of temperature control range is always 50
Note: Based on our measuring method.
86
CO
Incubator MCO-170AICUVHL
2
MCO-170AICUVHL-PA
o
Ambient temperature+5
C to 50
o
±0.25
C (ambient temperature: 23
o
±0.1
C (ambient temperature: 23
±0.15 % (ambient temperature: 23
Temperature: 5
(The designed performance may not be obtained
If the ambient temperature is equal or less than 15
AC 110 V-120 V, 60 Hz
o
C, temperature control range: 30
MCO-170AICUVL
MCO-170AICL
MCO-170AICUVL-PA
MCO-170AICL-PA
o
C* (ambient temperature: 5
o
o
C, setting: 37
o
o
C, setting: 37
C, CO
0 % to 20 %
o
o
C, setting: 37
C, CO
95 %R.H.±5 %R.H.
o
o
C to 35
C, Humidity: 80 %R.H. max.
29 dB (A scale)
Max. 376 W
Max. 1,070 kJ/h
Max. 3.4 A
o
o
C~50
C. Regardless of ambient
o
C.
o
o
C to 35
C)
C, CO
: 5 %, no load)
2
: 5 %, no load)
2
: 5 %, no load)
2
o
C)

Hide quick links:

Advertisement

loading

This manual is also suitable for:

Mco-170aicuvlMco-170aicuvhl