2
9. Set the humidifying pan cover, the humidifying pan removed in the procedure 1 (Fig.3)
Note: H
O
decontamination can be performed only for the chamber and inner attachments with standard
2
2
specifications, not for any other objects.
Front
10. Make sure that the duct, the fan and the H
page 82). After that, close the inner door and the outer door.
11. Press H
O
key for 3 seconds to lead the H
2
2
Decontamination screen.
Note:
・ When
the
model
decontamination
displayed on the Top screen.
・When key lock is ON, Password input window is
opened and input of the password is required.
Refer to page 34.
12. On the H
O
Decontamination Step1 screen,
2
2
the system check starts automatically. If the system
is
normal,
the
Decontamination Step2 screen. When the system
is abnormal, refer to Table 8 on page 79.
62
does
not
have
function,
H
O
key
is
2
2
display
leads
the
Humidifying
pan cover
Humidifying pan
H
O
2
2
generator
O
generator are securely attached (refer to Fig. 1 to 3 on
2
2
O
2
2
H
O
2
2
not
H
O
2
2
Fig. 3

Hide quick links:

Advertisement

loading

This manual is also suitable for:

Mco-170aicuvlMco-170aicuvhl