Combination operation
3.4 Plug & Measure: Permissible activities
Requirement for successful module retrofitting or successful module replacement
The firmware of the module to be replaced must be compatible with the firmware of the
processing module (PM).
Prior to the installation, you must update the firmware of the processing module:
● When installing a new module into an existing device.
● When replacing a module: When the error message 'Download firmware/configuration'
appears in the display.
If an analyzer module is already present, you must update its firmware.
See also
Plug & Measure: Permissible activities (Page 27)
Case differentiations (Page 30)
3.4.2

Setting instructions

NOTICE
Damage caused by inadmissible procedure/faulty operations
Inadmissible procedure and/or faulty operations can result in damage to the device or its
modules/electrical equipment in connection with Plug & Measure. Therefore:
• Observe the specifications regarding installation/removal of the device hardware.
• Carefully read the following information on different Plug & Measure cases.
NOTICE
Deletion of the analyzer module parameter assignment due to faulty operation during
commissioning
During commissioning after replacement of the AM electronics, a faulty operation can
prevent transfer of the parameters and render the AM electronics unusable. When you
commission the device once again after replacing the AM electronics, the following applies:
• First, confirm the message regarding change of the configuration
• Make sure to confirm the transfer of the parameters by pressing <ENTER>
28
Compact Operating Instructions, 06/2017, A5E31805656-05
Quick Start