Case Differentiations - Siemens SIPROCESS GA700 Quick Start Manual

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Combination operation
3.4 Plug & Measure: Permissible activities
3.4.3

Case differentiations

Table 3- 4
Case
PM
New
New
Data synchronization: PM parameters are transferred to the AMx (x=1 or 2), parameters of the
1)
AMx are transferred to the PM
Table 3- 5
Case
PM
Un-
changed
Un-
changed
Un-
changed
Un-
changed
Un-
changed
Data synchronization: PM parameters are transferred to the AMx (x=1 or 2), parameters of the
1)
AMx are transferred to the PM
30
Plug & Measure - Commissioning
Description
AMx
New
Commissioning, PM without serial
number
New
Commissioning, each module has a
serial number stored at the factory.
None of the modules has been con-
nected to another module before
Plug & Measure installation of an AM into an existing (configured) device / replacement
of an AM in an existing (configured) device
Description
AMx
Re-
The AM is to be replaced by a module
placed
of the same type. The replacement
AM has a different serial number than
the AM connected previously.
Replacement of the AM electronics,
only permitted without own serial
number stored in the AM electronics
Re-
AM electronics installed as "dedicated
placed
spare part". Service event.
Re-
An AM is replaced by an AM of a
placed
different type.
New
An additional AM of the same type is
to be installed.
An additional AM of a previously in-
stalled type is installed again.
New
An additional AM of a type that has
not been connected before is to be
installed.
Data synchronization/Action
Automatic data synchronization
Automatic data synchronization
Data synchronization/Action
User action. Confirm <ENTER> or
reject <ESC>.
The AM data is transmitted from the
PM to the AM after confirmation, ex-
cept the identification data and the
calibration parameters.
User action. Always confirm <ENTER>.
The AM data is transmitted from the
PM to the AM after confirmation, in-
cluding the identification data and the
calibration parameters.
Normal startup without message. The
default data of the AM set in the factory
is applied.
Automatic data synchronization
action required by user.
User action. Confirm <ENTER> or
reject <ESC>.
The data is transmitted after confirma-
tion, except the identification data and
the calibration parameters.
Automatic data synchronization
action required by user.
Compact Operating Instructions, 06/2017, A5E31805656-05
.