Measure the capacitor bank voltage on the bus bar (Fig. 3).
If one of the capacitor series of the capacitor bank is still charged (U > 10 V), it can be
discharged via resistor R103 on D7.
Connect the voltmeter to the measurement points parallel to the insulated cables which
will be used to discharge the capacitor bank (U < 10 V) (Fig. 4).
MOBILETT Plus / - E/ - M
SPR8-215.898.01
Rev. 05
01.02
Page 6 of 6
Siemens-Elema AB
SPS-UD
General
Solna, Sweden