Siemens 6 Series Compact Operating Instructions page 10

Continuous gas analysis
Table of Contents

Advertisement

Application planning
Harmful substances in the condensate
The condensate at the sample gas cooler or condensation trap can be corrosive, depending on
the composition of the sample gas (e.g.: through SO
paper into the condensate it will turn red.
• Wear protective clothing, protective goggles and protective gloves when handling the
• Prior to its disposal, dilute the condensate with tap water until it turns neutral and the litmus
Strong vibrations
Vibrations can influence the result of a measurement.
• The analyzer must therefore only be installed at a location which is free of vibration.
• Ensure that there is little vibration when using the device. The maximum vibrations must
NOTICE
Manipulation of the device
Physical access to the device enables undesired manipulations of the device. As a result of this,
the device can lose its measuring properties and become unusable.
• Ensure that the device is located in a protected area to which only authorized persons have
• If no settings need be carried out, lock the device.
Note
Measuring high oxygen concentrations
For the measurement of high O
all parts that come in contact with the sample gas are free of oil and grease and do not contain
any particles that could react with O
We highly recommend you use specially cleaned analyzer modules with the "Cleaned for O2"
option for such measurements.
10
WARNING
condensate!
paper does not change color. Alternatively, you can also neutralize the condensate with a
weak sodium carbonate solution (Na
CAUTION
not exceed 5 m/s².
access.
, NO, H
2
CO
).
2
3
concentrations in excess of 25%, it is absolutely necessary that
2
.
2
Compact Operating Instructions, 09/2018, A5E45779403002-01
S, HCl, etc.). When you dip litmus
2
Safety instructions