Philips PageWriter 100 Instructions For Use Manual page 26

Table of Contents

Advertisement

Recording an Auto ECG
2-4
5HFRUGLQJ DQ $XWR (&*
To record an Auto ECG, perform the following steps.
1. If the cardiograph is not On, press
lights are lit as the cardiograph performs a short power-on
sequence.
2. Prepare the patient and apply the electrodes.
$XWR
3. Press
on the front panel.
The light above the
acquiring the ECG. (If the
pattern with the
See "Correcting a Leads Off Condition" below.) When the
cardiograph begins printing the ECG, the
When the cardiograph has completed the report, the light turns off.
&RUUHFWLQJ D /HDGV 2II &RQGLWLRQ
A leads off condition is caused by the patient cable not being connected
to the cardiograph, one or more leadwires not being connected to
electrodes, or poor contact between the patient and one or more
electrodes. The cardiograph signals a leads off condition only during
the acquisition of an Auto ECG. The operator can check for a leads off
condition anytime an ECG report is printed by checking for traces
consisting of dashed lines or by checking the upper left corner of the
Auto ECG report for a leads off message. The leads off message lists
the lead wires or electrodes that have a leads off condition. Table 2-1
shows the possible leads off messages, and the lead wire and electrode
to check for each one. To correct a leads off condition, perform the
following steps.
1. Check that the patient cable is properly connected.
2. Recheck the steps you performed when preparing the patient. Pay
particular attention to the proper application of the electrodes and
the connections between leadwires and electrodes.
$XWR
key flashes while the cardiograph is
$XWR
light flashes in an alternating
0DQXDO
lights a leads off condition is indicated.
2Q6WDQGE\
. The front panel
light stays on.
$XWR
Recording an ECG

Advertisement

Table of Contents
loading

This manual is also suitable for:

M1772a

Table of Contents