Philips PageWriter 100 Instructions For Use Manual page 27

Table of Contents

Advertisement

PageWriter 100 M1772A Cardiograph
7DEOH  /HDGV 2II /DEHOV
Designator (AHA/IEC)
RL/N
RA/R
LA/L
LL/F
V1 ... V6/
C1 ... C 6
$XWR 5HSRUW )LOWHUV
The PageWriter 100 allows you to choose between two sets of filters
for Auto ECG reports:
l
The 0.15-40 Hz filters (0.15 Hz Baseline Wander and 40 Hz
Noise) are enabled when the Filter light is on. It delivers the
least amount or noise.
l
The 0.15-150 Hz filters (0.15 Hz Baseline Wander and 150 Hz
Noise) are enabled only when the Filter light is off. It delivers
the highest fidelity signal, but unless ECG signal conditions are
excellent, it can record noise.
Table 2-2 shows when each filter is enabled.
7DEOH  $XWR )LOWHU 6HWWLQJV
Default (Filter light on)
0.15-40 Hz
a. Conforms to AAMI specifications
Meaning
Right leg electrode not connected, or only right leg
electrode is connected and all other limb electrodes
are not connected.
Right arm electrode is not connected.
Left arm electrode is not connectd.
Left leg elecrode is not connected.
One or more chest electrodes are not connected. For
example, V2 means the V2 electrode is not con-
nected.
Filters Off (Filter light off)
a
0.15-150 Hz
Auto Report Filters