Checking The Shock Sensors - Siemens MOBILETT XP Digital Installation And Startup

Hide thumbs Also See for MOBILETT XP Digital:
Table of Contents

Advertisement

30
5-

Checking the shock sensors

4Checking the shock sensors
P
Prior to startup, the 4 shock sensors must be checked.
Fig. 12: Shock sensor
If a shock sensor has been activated (colored red), the CXDI detector has to be replaced
before start-up.
Check the shock sensors.
MOBILETT XP Digital
SPR8-230.814.30.03.02
11.05
Checking the shock sensors
Page 30 of 70
CS SD 24