30
5-
Checking the shock sensors
4Checking the shock sensors
P
Prior to startup, the 4 shock sensors must be checked.
Fig. 12: Shock sensor
If a shock sensor has been activated (colored red), the CXDI detector has to be replaced
before start-up.
•
Check the shock sensors.
MOBILETT XP Digital
SPR8-230.814.30.03.02
11.05
Checking the shock sensors
Page 30 of 70
CS SD 24
0
Siemens AG
Medical Solutions