Start-up report
10-
MOBILETT XP Digital
9Start-up report
This protocol confirms operability according to the datasheet. The original document
remains with the unit, and a copy is filed at the local branch office.
Type / Material No. / Serial No.: ...................................................................................
Customer: ............................................. Customer No.:
........................................................
Check
❐
❐
❐
❐
❐
❐
❐
❐
❐
❐
❐
❐
❐
❐
❐
Options
❐
❐
Quality checks
Siemens AG
Medical Solutions
Control point
Required documents complete
Visual inspection
Cable winch
Support arm movements
Single tank and collimator movements
Shock sensors
Line voltage
Battery charging operation
Exposure operation
Lead apron holder
Collimator function
Calibrating the CXDI detector
Exposure release with high voltage
Configuration of the imaging system
Light field/radiation field
DAP (test display 80 - 120 µGy m
µGym)
Remote control
SPR8-230.814.30.03.02
11.05
Additional information
_________________V
Deviation ≤ ± 1.7%
2
Display _________ µGym
or 8-12
Page 57 of 70
CS SD 24
57
0
2
MOBILETT XP Digital