General start-up information

Start-up information

The start-up procedure described here corresponds to a functional test. All adjust-
ments and calibrations are performed at the factory. The MOBILETT XP Digital is ready
for operation on completion of the start-up report.
The top cover has to be removed to access the CXDI PC or perform measurements on
D916. Place the top cover on a table or similar surface. All cables remain connected.
The unit can be connected to line voltages of 100-130 V ( ± 10%) or 200-240 V ( ±
10%). The unit automatically adjusts to the existing line voltage.
NOTE
Following start-up, country-specific tests must be performed if necessary.
For example:
Acceptance test according to §16 of the X-ray Ordinance (Germany):
Use the measurements in the test certificate provided as the initial values for the required
measurements.
The following values were measured by the manufacturer; refer to the test certificate:
- Brightness of the light localizer
- Filter values (Al equivalent for tube assembly and collimator)
- Coincidence of light field and radiation field
- Accuracy of the tube voltage
- Accuracy of the kV value steps
- Accuracy of the mAs values
- Reproducibility of dose values
The test/measured values marked by the report icon must be entered in the start-up re-
port located at the end of this manual.
NOTE
Siemens AG
Medical Solutions
Observe the nominal current of the fuses for the on-site power
connections:
- 15A slow-blow (on-site) for 100-130V ± 10% or
- 15A slow-blow (on-site) for 200-240 V ± 10% line voltage.
The power cable for this device is equipped with a standard safety
plug.
Attach a power plug that is compliant with local standards if nec-
essary.
Recording information immediately during start-up saves time.
Fill in the report form after completing each task.
The start-up report is a component of this documentation.
SPR8-230.814.30.03.02
11.05
MOBILETT XP Digital