BEFORE COMMENCING OPERATION

Preparation of monitor bottle
Prior to the operation of the unit, fill the monitor
bottles (upper and lower) with 10% glycerol or
water using the following procedure:
1. Remove the top and bottom drawer (or shelf).
2. Remove the monitor bottle cover as shown in
the figure 1.
3. Take out the sensor from the monitor bottle.
4. Remove the fixing screw and take out the
monitor bottle.
5. Remove the cap by loosing it as shown in the
figure 2.
6. Pour 200ml of 10% glycerol or water into the
monitor bottle.
7. Cap the monitor bottle and replace the bottle to
its original position. Then set the sensor to the
bottle and replace the cover on the bottle.
Recorder sensor
(lower bottle of MBR-304G
and MBR-304GR)
Fixing screw
Cap
13
Temp. sensor
Cover
Fig. 1
Loosen
Fig. 2