Remote Alarm Terminal - Sanyo BLOOD BANK MBR-304DR Instruction Manual

Blood bank refrigerator
Table of Contents

Advertisement

BEFORE COMMENCING OPERATION

Remote alarm terminal

The terminal for remote alarm is installed inside the rear
cover of the unit.
The alarm is outputted from this
terminal.
The contact rating is DC 30V, 2 A.
Contact output: At normal condition "Open"
At abnormal condition "Close"
Note:
The contact is "Close" because it is regarded as
abnormal when the power supply cord is disconnected
Terminal
from the outlet or the power switch is turned off, or in
the event of power failure.
The remote alarm is in conjunction with the alarm
buzzer, the alarm can be silenced by pressing the alarm
buzzer stop key (BUZZER) except for in the case of
power failure
14