OPERATING INSTRUCTIONS

Storing of items
Do not disturb the air circulation in the chamber by the stored items. The disturbed air circulation can
cause the freezing of stored items or degrade of temperature distribution in the chamber.
Air-intake vent
Do not block the air-intake by the stored items near the air-intake vent. Take should be taken not to put
any paper or plastic chips near the vent. They may be vacuumed up and block the vent.
Air-exhaust vent
Do not block the air-exhaust by the stored item near the air-exhaust vent. The stored item may be
frozen if they put nearby the air-exhaust vent.
Load line
The stock area is limited to ensure the air flow in the chamber. The labels "load line" are attached to the
right and left side of the chamber. Always put the items within the specified area.
Note:
Do not stock any corrosive materials when the container is not sealed completely. Such material causes
corrosion of inner components or evaporator.