Siemens POLYMOBIL Plus Start-Up Instructions
Hide thumbs Also See for POLYMOBIL Plus:

Advertisement

Quick Links

POLYMOBIL Plus
Start-up Instructions
Print No.: SPR8-125.034.01.03.02
Replaces: SPR8-125.034.01.02.02
SP
© Siemens AG 1998
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model
_or_ design,_are_
reserved.
English
Doc. Gen. Date: 01.00

Advertisement

Table of Contents
loading

Summary of Contents for Siemens POLYMOBIL Plus

  • Page 1 POLYMOBIL Plus Start-up Instructions © Siemens AG 1998 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility...
  • Page 2 0 - 2 Revision Chapter Page Revision POLYMOBIL Plus SPR8-125.034.01 Page 2 of 4 Siemens AG Rev. 03 01.00 TD SD 24 Medical Engineering...
  • Page 3: Table Of Contents

    5 _______Test protocol __________________________________________________ 5 - 1 POLYMOBIL Plus Start-up protocol ......5 - 1...
  • Page 4 0 - 4 Contents This page intentionally left blank. POLYMOBIL Plus SPR8-125.034.01 Page 4 of 4 Siemens AG Rev. 03 01.00 TD SD 24 Medical Engineering...
  • Page 5: Specially Marked Text

    ARTD / Part 3. • Standard service kit • Protective conductor test meter • Equivalent leakage current meter • Digital multimeter. Siemens AG SPR8-125.034.01 Page 1 of 6 POLYMOBIL Plus Medical Engineering Rev. 03 01.00 TD SD 24...
  • Page 6: Safety Information And Protective Measures

    When performing service work and tests, adhere to: CAUTION - the product-specific safety information in the document, - the safety information in RA0-000.012..in the Register of the POLYMOBIL Plus binder, - as well as the general safety information contained in part 2 of the ARTD. •...
  • Page 7 This can cause life-threatening voltage to be present in the system even after a prolonged period of time. • Refer also to the POLYMOBIL Plus Service instructions RXR8-120.061.01... "Replacing the capacitor bank". Siemens AG SPR8-125.034.01 Page 3 of 6...
  • Page 8: Important Start-Up Information

    After completion of the start-up procedure, the POLYMOBIL Plus which has been completely tested and adjusted at the factory, is ready to operate. • The POLYMOBIL Plus can be connected to nominal line voltages from 110 V AC through ± 230 V AC ( 10%) and automatically self-adjusts to the existing power line voltage.
  • Page 9: Replacing Damaged Or Missing Screws

    Do not use abrasive cleaning agents or organic solvents such as benzine, alcohol or spot remover. • Do not spray water on the unit. • For additional information, refer to the Operating Instructions "Cleaning/Disinfecting". Siemens AG SPR8-125.034.01 Page 5 of 6 POLYMOBIL Plus Medical Engineering Rev. 03 01.00 TD SD 24...
  • Page 10 1 - 6 Requirements This page intentionally left blank. POLYMOBIL Plus SPR8-125.034.01 Page 6 of 6 Siemens AG Rev. 03 01.00 TD SD 24 Medical Engineering...
  • Page 11 Place the side panel with the integrated ramp (1/Fig. 1) down directly in front of the pallet and engage the retaining tabs. • Remove the plastic cover. Siemens AG SPR8-125.034.01 Page 1 of 2 POLYMOBIL Plus Medical Engineering Rev. 03 01.00 TD SD 24...
  • Page 12 Remove the 2 Allen screws and the bracket (7/Fig. 2, 17 mm spanner opening) from the stand trolley on the right and left side. • Push the POLYMOBIL Plus over the rail (8/Fig.3) and move it off the pallet via the ramp (9/Fig. 3). POLYMOBIL Plus SPR8-125.034.01...
  • Page 13: Checking The Nominal Line Voltage

    • Attach the warning label printed in the appropriate language, item number 18 13 554, to the operating console (1/Fig.1). NOTICE The warning label is located in the POLYMOBIL Plus binder. Fig. 1 Siemens AG SPR8-125.034.01 Page 1 of 6...
  • Page 14: Checking The Controls And Indicators

    Press the ON key (5/Fig. 5) on the control console. ➪ After switching the system on, it executes an initialization routine as follows: Initialization The POLYMOBIL Plus is ready – – – – – – for exposure: the kV display no...
  • Page 15: Checking The Counterbalance System

    Siemens AG SPR8-125.034.01 Page 3 of 6 POLYMOBIL Plus Medical Engineering Rev. 03 01.00 TD SD 24...
  • Page 16: Checking The Controls On The Double-Slot Diaphragm

    Close the double-slot diaphragm. Set 60 kV, 10 mAs. • Release an exposure. The radiation indicator (9/Fig. 5) lights during an exposure; at the same time, an acoustic signal sounds. POLYMOBIL Plus SPR8-125.034.01 Page 4 of 6 Siemens AG Rev. 03 01.00...
  • Page 17: Checking The Exposure Control

    On the developed film, write the following data with a waterproof felt pen: - SID setting - film size - radiation field size Siemens AG SPR8-125.034.01 Page 5 of 6 POLYMOBIL Plus Medical Engineering Rev. 03 01.00 TD SD 24...
  • Page 18 NOTICE In case of deviations greater than 1.6 cm, see Service Instructions Chapter "Checking coincidence of light and radiation fields". • Switch OFF the POLYMOBIL Plus and disconnect the line voltage cable. POLYMOBIL Plus SPR8-125.034.01 Page 6 of 6 Siemens AG Rev.
  • Page 19: Protective Conductor Measurement

    The initial value measured may not exceed the maximum value of 2 mA. Conform to the measurement setup shown in Fig. 2. During the measurement, POLYMOBIL Plus must be switched on. Therefore, when using the BENDER Safety tester, it must be set to manual measurement.
  • Page 20 (L) N 1-Unit connection lead (without protective conduc- contact) 2-Patient connection lead (not on POLYMOBIL Plus) 3-System housing 4- Measurement circuit Fig. 2 - Measuring the equivalent leakage current for systems in operation. • Roll up the power line cable on the cable holder in back of the control console.
  • Page 21: Polymobil Plus Start-Up Protocol

    Test protocol 5 - 1 POLYMOBIL Plus Start-up protocol The Start-up protocol follows the same format as the start-up instructions. Item number / Type number / Serial number: ....
  • Page 22 5 - 2 Test protocol This page intentionally left blank. POLYMOBIL Plus SPR8-125.034.01 Page 2 of 2 Siemens AG Rev. 03 01.00 TD SD 24 Medical Engineering...
  • Page 23 6 - 1 Page 1-3: ➪ "350 V" changed into "450 V". TD SD 24 / Am-Ende TD SD 24/ Groß SMS Iselin / O’Donnell Siemens AG SPR8-125.034.01 Page 1 of 2 POLYMOBIL Plus Medical Engineering Rev. 03 01.00 TD SD 24...
  • Page 24 6 - 2 Changes to previous version This page intentionally left blank. POLYMOBIL Plus SPR8-125.034.01 Page 2 of 2 Siemens AG Rev. 03 01.00 TD SD 24 Medical Engineering...