Advertisement

Siremobil Compact L
Replacement Instructions
15" TFT Display 71 42 354
Print No.: SPR2-130.841.03.02.02
Replaces: SPR2-130.841.03.01.02
SP
© Siemens AG 2005
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model
_or_ design,_are_
reserved.
English
Doc. Gen. Date: 06.05

Advertisement

Table of Contents
loading

Summary of Contents for Siemens Siremobil Compact L

  • Page 1 Siremobil Compact L Replacement Instructions 15" TFT Display 71 42 354 © Siemens AG 2005 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights,...
  • Page 2: Document Revision Level

    Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
  • Page 3: Notes And Symbols

    Siemens AG SPR2-130.841.03 Page 3 of 20 Siremobil Compact L Medical Solutions Rev. 02 06.05 CS PS 24...
  • Page 4: Validity Of These Instructions

    230V ~ Power cable not included. External switched-mode power 07141083 230V ~ Power cable not included. supply Display holder 07142347 Fastening screws not included Siremobil Compact L SPR2-130.841.03 Page 4 of 20 Siemens AG Rev. 02 06.05 CS PS 24 Medical Solutions...
  • Page 5: Safety Information

    5 minutes before resuming work. CAUTION Electrical voltage! Noncompliance can lead to property damage. ESD protection guidelines must be observed when servicing. Siemens AG SPR2-130.841.03 Page 5 of 20 Siremobil Compact L Medical Solutions Rev. 02 06.05 CS PS 24...
  • Page 6 NOTE (permanent visible OSD menu) by selecting the OSD menu "Options" - "Onscreen Menu" - "OSD Time". After the adjustments, set it back to the default value (20 seconds) Siremobil Compact L SPR2-130.841.03 Page 6 of 20 Siemens AG Rev. 02 06.05...
  • Page 7: Troubleshooting

    3. Position to the front: BNC input selected, 75 Ohm termination resistor is on. - Check the settings of the on screen menu, in accordance with the user manual. Siemens AG SPR2-130.841.03 Page 7 of 20 Siremobil Compact L Medical Solutions Rev. 02 06.05 CS PS 24...
  • Page 8 Install the display(s) on the holder (1 / Fig. 1). • Reconnect the low voltage connector and video signal connector to each TFT display. • Check the functions of the TFT displays. Siremobil Compact L SPR2-130.841.03 Page 8 of 20 Siemens AG Rev. 02 06.05...
  • Page 9 Fix the low voltage cable with cable ties. • Connect the power plug to the switched-mode power supply. • Check the function of the TFT display. Siemens AG SPR2-130.841.03 Page 9 of 20 Siremobil Compact L Medical Solutions Rev. 02 06.05 CS PS 24...
  • Page 10 • Check the automatically detected video mode. • Check the geometry of the displayed image. • Adjust the brightness, contrast and back light intensity. Siremobil Compact L SPR2-130.841.03 Page 10 of 20 Siemens AG Rev. 02 06.05 CS PS 24...
  • Page 11 1 Current 0 Current Tab. 1 • If the values are different to table Tab.1, follow the description "Manual programming" (page 16). Siemens AG SPR2-130.841.03 Page 11 of 20 Siremobil Compact L Medical Solutions Rev. 02 06.05 CS PS 24...
  • Page 12 • Check the geometry of the displayed test pattern. It should be displayed as shown in Fig. 4. If adjustments are necessary, follow the description "Manual programming" (page 16). Siremobil Compact L SPR2-130.841.03 Page 12 of 20 Siemens AG Rev. 02 06.05...
  • Page 13 "DOWN" key and select it with the "Enter" key. • Set Brightness, Contrast and Back light intensity to the middle value "50%" and confirm it by pressing the "Enter" key. Siemens AG SPR2-130.841.03 Page 13 of 20 Siremobil Compact L Medical Solutions Rev. 02 06.05 CS PS 24...
  • Page 14 (OSD menu "Options" - "Onscreen Menu" - "OSD Time"). Final work steps Close all covers, if opened. Perform the Protective conductor test in accordance to the ARTD002.731.17... The protective conductor resistance may not exceed 0,2 Ohms. Siremobil Compact L SPR2-130.841.03 Page 14 of 20 Siemens AG Rev. 02 06.05...
  • Page 15: Manual Programming

    Scroll to line "back" with the "UP" and "DOWN" key and select it by pressing the "Enter" key. The main menu is displayed. Siemens AG SPR2-130.841.03 Page 15 of 20 Siremobil Compact L Medical Solutions Rev. 02 06.05 CS PS 24...
  • Page 16 "Enter" key. • Scroll to line "back" with the "UP" and "DOWN" key and select it by pressing the "Enter" key. The main menu is displayed. Siremobil Compact L SPR2-130.841.03 Page 16 of 20 Siemens AG Rev. 02 06.05...
  • Page 17 Scroll to line "back" with the "UP" and "DOWN" key and select it by pressing the "Enter" key. The main menu is displayed. Siemens AG SPR2-130.841.03 Page 17 of 20 Siremobil Compact L Medical Solutions Rev. 02 06.05 CS PS 24...
  • Page 18: Horizontal And Vertical Position

    Confirm the adjusted value by pressing the "Enter" key. • Scroll to line "back" with the "UP" and "DOWN" key and select it by pressing the "Enter" key. The main menu is displayed. Siremobil Compact L SPR2-130.841.03 Page 18 of 20 Siemens AG Rev. 02 06.05...
  • Page 19 Check the brightness, contrast and back light intensity as described in section "Brightness, contrast and back light" • Perform the work steps as described in section "Final work steps". Siemens AG SPR2-130.841.03 Page 19 of 20 Siremobil Compact L Medical Solutions Rev. 02 06.05 CS PS 24...
  • Page 20 Page 11, table Tab1, line "Mode": In column 50Hz, value changed to 614x514 Page 17, value "Picture Width" for 50Hz video frequency changed to 614. Page 17, value "Picture Height" for 50Hz video frequency changed to 514. Siremobil Compact L SPR2-130.841.03 Page 20 of 20 Siemens AG Rev.

Table of Contents