Function checks
1
2
Fig. 2

Checking the counterbalance system

Unlock the support arm (1/Fig. 2) and loosen the knob (2/Fig. 2). The support arm
without attached accessories should move through the entire range easily and should
stop in any position.
Readjusting the counterbalance system with accessories
attached
If the support arm with accessories attached does not stop in any desired position,
readjust the spring tension for the counterbalance.
Remove the cover (3/Fig. 2) from the stand.
Release the supplemental brake on the support arm (2/Fig. 2).
Move the support arm into the horizontal position.
Tighten the screw (4/Fig. 3) using a 17-mm open-ended wrench. Adjust the spring
tension in such a way, that there is a balance when the support arm is in the horizontal
position, i.e. that equal force is required to lift and lower the tube assembly.
Siemens AG
Medical Solutions
3
Fig. 3
SPR8-125.034.02
Rev. 02
03.05
4
Page 3 of 6
CS PS 24
3 - 3
3
3
POLYMOBIL Plus