3 Column stand

SIM
C-Arm carriage wheels
Do the wheels show any damage or sign of wear (n°1, see figure)?
SIM
C-Arm carriage rails
Do the rails show any damage or sign of wear or loosened fixation (n°2, see figure)?
SIM
Gas spring
Does the gas spring assembly show any damage or sign of wear or loosened fixation (n°3, see
figure)?
Does the alarm come out if safety micro-switch is disconnected (n°4, see figure)?
SIM
Vertical C-Arm drive
Do the chain, chain fixation and gear box show any damage or sign of wear or loosened fixations
(n°5, see figure)?
SIE
Limit switches
System ON. Move the lifting carriage up and down.
Do the first limit switches MW4 (up) and MW3 (down) stop the carriage travel at the top and at the
bottom?
Manually activate safety switches MW2 (up) and MW1 (down) when carriage is at safety distance.
Lifting carriage must be blocked.
PMF
Vertical travel
Is the vertical travel smooth and noiseless?
MAMMOMAT Balance
Maintenance Instructions
SPB7-115.831.01
Page 10 of 22
Rev.01
05.04
3 Column stand