5 Rotating anode starter

PMF Protection circuit check
For protection circuit check, the anode must be put in rotation and it is advisable as follows:
DEMO mode with DIP 3 SW1 of board 01-170 switched ON
Manual mode with at least 200 mAs
In this mode, no X-rays are generated and measurement time is sufficiently long.
PMF Control LED (brake enabled)
PCB 92-074
LED1
LED2
LED3
LED5
LED6
MAMMOMAT Balance
Maintenance Instructions
Stand by
Stand by
(yellow) bright during normal rotation phase (RUN)
(yellow) bright during breaking or alarm
(yellow) bright during breaking
(green) bright in stand by or rotation
(green) bright during rotation
SPB7-115.831.01
Rev.01
5 Rotating anode starter
V(TP3) = 0.65xV(TP1)
V(TP4) = 0.65xV (TP2)